Fit Issues upon Delivery

Sometimes the first aligner may not seat correctly, this could be for numerous reasons:

  • Impressions/scanning issues – if there is a discrepancy between the records provided and the patients teeth, this will cause the aligner to not seat fully. This could either be due to a distortion, false information, or the patients’ teeth have moved since the original records were provided.
  • Path of insertion – If a patient has not worn aligners or retainers before they may not be putting them in correctly. When the patient has severe crowding, it can be difficult to seat the aligners, and this may take several days to get used to.
  • Production error – On rare occasions, there may have been a manufacturing error at the laboratory which has caused the aligners to not seat. This could be within the printing or the manufacturing process of the aligners. To eliminate this, our aligners are all manufactured by hand by our experienced technicians with rigorous quality and accuracy procedures in place.

What to do Next

We understand it can be disheartening when the patient’s first aligner does not seat. The first action would be to help the patient seat the aligner correctly; this could be by using chewies for a few seconds starting at the molars and moving towards the centrals. Having a small gap incisally when seating an aligner for the first time is quite common, as we are trying to move the teeth into a different position. If the aligner is close to seating, it may be suitable to send the patient away for a few days to see if the aligner seats more comfortably as the teeth begin to move. The majority of cases experiencing fitting issues will be resolved after

We advise contacting the Smilelign team directly for advice if the first steps do not work, as they can advise how to proceed. The technicians will likely request a new scan/impression if the aligners do not seat. Once the lab receives these, they will be assessed to find the reason why the aligner did not seat, they can then proceed to replan using the new records.

Tracking Issues

It is not uncommon for clear aligner patients to experience tracking issues at some point in treatment. It is evident when the patient is not tracking as planned as the aligner will not engage with the attachments correctly and there are visible gaps incisally. The most common reasons for tracking issues are patient compliance, and uneven tooth movement.

Patient compliance is huge factor in the success of clear aligner treatment. It is recommended that the patient wear the aligners for no less than 22 hours a day, only removing them to eat or clean. Any deviation from this recommendation can cause the teeth to not progress as planned and therefore make the aligners not fully seat. Uneven tooth movement is seen in some instances where some teeth move as planned and others do not. Each tooth can have a different response to the orthodontic movement dependent on the root, bone and crowding.

Similarly, if you feel a case is not tracking, we advise you to send in a new impressions or scans for assessment. The technicians can overlay the predicted position with the current position and assess which teeth are not tracking and assess why there is a tracking issue. At this stage the technician would replan the case from the new record and provide further aligners, advising you or any additional IPR/attachments.

Fitting and tracking issues with clear aligners do happen from time to time, though they’re not frequent. If a patient experiences any difficulty, please don’t hesitate to contact us. Our friendly and experienced team is here to advise you on the best solution to get things back on track quickly.

Start offering aligner treatment sooner, with expert support every step

You don’t need to book a course or invest in new equipment to offer Smilelign. Just send your first case and we’ll handle the rest.

  • No training or subscriptions required
  • We accept all major scanners (plus impressions)
  • Free treatment plan with your first case
  • Support from day one via call, chat or email
  • Quick-start guide & Academy access included

Aligners that grow your bottom line not your workload

Smilelign is more than a clinical solution… It’s a business tool. We help you convert more patients, streamline your consultations and increase revenue per treatment. With bundled whitening, retention and reward incentives, there’s never been a better time to start

  • Conversion tools + scripts included
  • Whitening + Retention bundle options
  • Reward scheme based per practice
  • Vitalisse discounts via Smilelign rewards